Welch Allyn monitor service utility Overview The Welch Allyn Monitor Service Utility is a software tool for use, with the VSM 300 Series service manual, by biomedical technicians, clinical engineers, and other qualified service personnel, to update and repair VSM 300 Series monitors. Factory or at an authorized Welch Allyn service center. Related documents When using this manual, refer to the following:. Welch Allyn Connex Vital Signs Monitor 6000 Series Directions for use (part number 103501). Welch Allyn Service Tool CD (part number 103521). Welch Allyn Service Tool Install guide (part number 103820).

Advertising

1

English

Set the date and time

After recharging a dead battery or after disconnecting the battery for a few minutes, you
must program the date and time screen.

You can change several device operating parameters in the Internal Configuration Mode.
When changed, these settings become the default power-up settings. You will also see
non-changeable device configurations for technical service purposes.

To enter the Internal Configuration Mode:

1.

Turn the Spot Vital Signs off (

).

2.

Press and hold the Blood Pressure Start/Stop (

) and Power (

) buttons. The

device enters the Internal Configuration Mode and displays the software version.

If the device displays E38, press the Next Patient/Clear/Cancel (

) button to

cancel the error.

3.

Press the Mode (

) button to cycle through the Internal Configuration menu until

you see the date option displayed on the screen.

4.

Use the Next Patient/Clear/Cancel (

) or Blood Pressure Start/Stop (

)

buttons to change the default setting.

Advertising
Table of contents
  • Chinese
    • 简体中文
Popular Brands
Popular manuals
Advertising

2

English

Welch Allyn Spot Vital Signs

5.

Press the Mode (

) button once to save the change. Repeat steps 4 and 5 to set

the day, month, year, hour, and minute values.

Allyn

6.

Press the Power (

) button to exit the Internal Configuration Mode.

Configuration Menu Options

Setting

Description

Blood Pressure Calibration
Displays “Cal”

Prepares the Spot Vital Signs for calibration. Only qualified personnel should verify the
Spot Vital Signs blood pressure calibration.

Inflation Pressure Preset Level
Displays “PrP”

120, 140, 160, 180, 200, 240, 280 mmHg. Factory default is 160 mmHg.

Pressure Preset Level
Displays “PrP“

On or off. Disables or enables the front panel Pressure Preset button.

Backlight
Displays “BLT”

On or off.

Welch Allyn Spot Vital Signs Monitor Manual

Mean Arterial Pressure
Displays “MAP”

On or off.

Welch Allyn Spot Vital Signs Monitor User Manual

Date/Time

Changes or updates the actual date and time.

Welch Allyn Vital Signs Monitor Manual

Temperature Scale
Displays “TMP MOD”

Fahrenheit (°F) or Celsius (°C) Normal Mode / Fahrenheit (°F) or Celsius (°C) Monitor
Mode

Welch Allyn Spot Vital Signs Troubleshooting

Blood Pressure Units
Displays “BP”

mmHg or kPa.

Welch Allyn Vital Spot Monitor

Battery Readings
Displays “BAT”

Displays the total battery voltage.

Battery Life
Displays “LFE”

Total number unit measurements. Displayed information only; operator cannot change.